จำหน่าย SHIMADZU UV-1900i Plus | UV-Vis-NIR Spectrophotometers

หมวดหมู่: แบรนด์:

คำอธิบาย

 

Refined User-Friendliness

 

Refined User-Friendliness

Easy-to-Use Interface
Grasp the Current Status and Operating Procedures at a Glance

  • The UV-1900i Plus on-screen user interface includes large, easy-to-see icons deployed on a black background, so the instrument settings are evident at a glance. In addition, the large, easy-to-see icons improve intuitive understanding, which enables users to quickly become familiar with the operations. Furthermore, the user interface is designed to minimize transitions between windows, so users do not get confused during the operations. The system also features a faster response speed by adopting a new CPU.
    Easy-to-Use Interface

    Display languages are available in eight languages (Japanese, English,Chinese, Spanish, Portuguese, German, French, Russian)

     

Multiple Measurement Modes

UV-1900i Plus has six basic measurement modes, which offer the optimal solution for a variety of measurements.                    

Multiple Measurement Modes

* The background color of the operation screen can also be set to white.

 

Various Functions for Comfortable Daily Measurement

Assist Function

  • Assist Function  AN

     The instrument assists the user to ensure that measurements are performed using the correct procedure.*1

Assist Function                               Assist Function                                      Assist Function
  • Checks whether or not baseline correction,                    – Informs the user if the last correction performed                   – Informs the user if the UV-1900i Plus is not ready to begin
    auto zero correction or cell blank correction                     is not appropriate for the planned measurement.                            measurement and when starting measurements and
    has been performed and informs the user                                                                                                                                                                     100 %T (0 Abs) corrections.
    if it has not been.

*1 This function can be enabled/disabled.

 

Shutdown Function

The shutdown function puts UV-1900i Plus to sleep after it’s been used, or after a certain period of time.
Putting the instrument into sleep mode limits power consumption and helps preserve the lamp.

Shutdown Function

Shutdown Function

 

Wakeup Function

Automatically wakes up a sleeping UV-1900i Plus at a specified time.

This function eliminates the need for analysts to wait for it to warm up,

allowing them to be more productive.

Wakeup Function

Startup Validation Function

Instrument performance checks can be performed automatically when the instrument is turned ON.*2
This enables even more reliable instrument operation.Startup Validation FunctionIt can also be implemented in combination with the Wakeup function.

*2 This function can be enabled/disabled.
The performance check items that can be used with this function are as follows: wavelength accuracy (D2), wavelength repeat accuracy (D2), degradation, noise level, baseline flatness, baseline stability (drift), [EP] Control of Wavelength Accuracy (D2), and [USP] Control of Wavelengths (D2).

 

Network Connectivity Function

Data can now be transferred to a PC via a network. With wireless printing, multiple UV units can print from a single printer.*3 (A router and other network equipment must be installed to use a network.)

Network Connectivity Function

 

*3 A router with a wireless function is required.
*4 Optional expanded memory is required. The instrument is not compatible with control via a network.
*5 A PictBridge-compatible printer is required.

 

Bar-Code Reader and Keyboard Entry Function

Sample names and numerical values can be entered by a bar-code reader or from the keyboard.
This saves time when entering sample names for a multiple-sample analysis, and prevents sample misidentification and other human errors.*6

Bar-Code Reader and Keyboard Entry Function

 

*6 Use a bar-code reader and keyboard with a USB connection.

 

High Performance to Meet Diverse Needs

 

Ultra-Fast Scan

Spectra can be acquired as fast as 29,000 nm/min. Ultra-fast scanning is effective in tracking chemical reactions in a short time. In addition to the absorbance change at specified wavelengths, spectra can also be acquired in a short time with the UV-1900i Plus. Therefore, more detailed behavior can be investigated by observing spectra with the UV-1900i Plus.

The figures below show the analysis of the particle agglomeration process when salts are added to silver nanoparticles. Measurements of the 300 to 700 nm region were performed in ultra-fast scan mode. In addition to the decrease of absorbance at 400 nm and the increase of absorbance at 480 nm, the temporal changes of spectra can also be observed.

Ultra-Fast Scan

Low Stray Light

Linearity will be lost in the high absorbance region due to stray light. UV-1900i Plus is equipped with Low Stray Light Diffraction Gratings, achieve “best-in-class” low stray light. Stray light is at 0.5 % max. (198 nm), making accurate measurements possible at 1 Abs or more, even in the ultraviolet region. In addition, high-concentration samples can be quantified accurately.
The figure on the right is a calibration curve for acetic acid, created with absorbance at 200 nm. The correlation coefficient is 0.9998 and correct measured values are obtained even with high-concentration samples.

Low Stray Light

 

High Reproducibility and Repeatability Accuracy

The photometric repeatability accuracy is 0.0002 Abs max. (0.5 Abs and 1.0 Abs). As a result, variance in the measurement results is suppressed, enabling more accurate quantitation and the detection of low-concentration samples.
The figure on the right is a calibration curve for caffeine, created with absorbance at 273 nm.
The calibration curve has an Abs = 0.0528 Conc. The lower limit of quantitation determined from the standard deviation is 0.0051 mg/L.

Note: One method of determining the lower limit of quantitation is to use ten times the standard deviation. This is an actual measured value and is not guaranteed.